Customization: | Available |
---|---|
After-sales Service: | Online Technical Support |
Application: | Laboratory Apparatus |
Still deciding? Get samples of US$ 1967/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Laboratory Single Beam Spectrophotometer Uv Visible Spectrophotometer With Good Price Range 190-1100 nm
Features
*High resolution C-T structure monochromator, good stability, reproducibility and accuracy.
*Advanced ARM microcomputer technology, excellent electronic circuit and signal processing, stable and reliable data, U disk data storage and export, one-key software upgrade, Bluetooth printing and other data output.
*Integrated aviation-grade aluminum alloy substrate and machined mirror base to ensure long-term stability of the optical path.
Product Characteristic
1. Photometric measurement: in the range of 190-1100nm, you can select the single-point test wavelength and test method you need to measure the absorbance or transmittance of the sample. The concentration of the sample can also be read directly by entering the standard concentration or concentration factor.
2. Quantitative measurement: measure the sample solution of unknown concentration through the curve of known parameter factors or automatically establish the standard solution curve; it has first-order \first-order zero-crossing, second-order, third-order curve fitting, and single-wavelength calibration, dual Wavelength calibration and three-point method are optional; standard curve can be stored and recalled.
3. Qualitative measurement: Set a wavelength range, scan interval, and then measure the absorbance, transmittance, reflectivity and energy of solid or liquid samples at intervals. It is also possible to superimpose, zoom, smooth, filter, detect peaks and valleys, save, and print the measured maps.
4. Time measurement: Time measurement is also called kinetic measurement. The sample is scanned according to the time range interval of absorbance or transmittance at the set wavelength point, and the absorbance can also be converted into concentration or reaction rate calculation by inputting the concentration factor.
Enzyme kinetic reaction rate calculation. You can also choose from a variety of spectrum processing methods such as stacking, scaling, smoothing, filtering, peak-valley detection, and derivation of the measured spectrum.
*Integrated aviation-grade aluminum alloy substrate and machined mirror base to ensure long-term stability of the optical path.
7-inch color large-screen graphic LCD display, with the navigation menu interface design, clear operation commands, with special functions to prevent users from operating errors, users can operate without reading the manual; double operation mode of touch screen + button, with gloves can operate.
*The four-slot 10-50mm universal adjustable manual sample holder is convenient for testing cuvettes of different specifications.
*Independently designed replacement window for Deut. lamp and tungsten lamp, flange type positioning structure, easy to replace.
*Strict manufacturing process and 32 key quality control points make each instrument very durable.
*Large-capacity memory, can store massive test data.
Model | U-6 |
Optical system | Single beam; 1200/mm imported holographic grating |
Wavelength range (nm) | 190-1100 |
Spectral Bandwidth (nm) | 2 |
Maximum allowable error of wavelength (nm) | ±0.5 |
Wavelength repeatability (nm) | ≤0.1 |
Maximum allowable error of transmittance (%T) | ±0.3%T; ±0.002Abs (0-0.5Abs); ±0.004Abs (0.5-1.0Abs) |
Transmittance Repeatability (%T) | ±0.1%T; ±0.001Abs (0-0.5Abs); ±0.002Abs (0.5-1.0Abs) |
%0 line and 100% line noise (%T) | 0% line 0.2/100% line 0.3 |
Drift (%T) | ≤0.2 |
Baseline Flatness (Abs) | ≤0.001 |
Baseline dark noise (%T) | ±0.3 |
Stray Light (%T) | ≤0.03@220nm and 360nm |
Supply voltage adaptability (%T) | ≤0.2 |
Photometric range | 0-200%T, -0.301-3A, 0-9999C(0-9999F) |
Test mode | Transmission, Reflection, Absorbance, Concentration, Energy |
Light source | Deut lamp + tungsten lamp |
Show | 7-inch touch screen display (resolution 800*480) |
System language | Chinese, English, Russian, Spanish |
Software upgrade | Support one-key upgrade of U disk |
Data output | U disk, Bluetooth, USB software upload |
Data output format | CSV |
Operating mode | Touch screen |
Sample holder | Four-slot 10-50mm Universal Adjustable Manual Sample Holder |
Host function | Basic + quantitative + qualitative + time + DNA + multi-wavelength |
Printer | Optional, support Bluetooth printing |
Power range | AC90~250V/ 50~60Hz |
Dimensions (L×W×H)mm | 470×330×210 |
Weight | 12kg |
T9 series touch screen dual beam UV-visible spectrophotometer
model | U-T9 | U-T9S | |
Optical system | Czerny-Turner type double beam; 1200 pieces / mm imported holographic grating | ||
Wavelength range | 190~1100nm | ||
Spectral bandwidth | 1nm | 0.5/1/2/4/5nm(Five-speed continuous adjustable) | |
Wavelength accuracy | ±0.3nm | ||
Wavelength repeatability | ±0.1nm | ||
Transmittance accuracy | ±0.3%T | ||
Transmittance repeatability | ±0.1%T | ||
Stray light | ≤0.03%T, @220nm/360nm | ||
noise | 0% line noise: 0.05%; 100 line noise: 0.1% | ||
drift | ±0.0008Abs(Preheat for more than 1 hour at 250 and 500 nm) | ||
Baseline flatness | ±0.0008Abs(Warm up for more than 1 hour) | ||
Baseline dark noise | 0.05%T | ||
Photometric range | 0~200℅T,-4~4A,0~9999C(0-9999F) | ||
Test mode | Absorbance, transmittance, reflectivity, energy | ||
light source | mported tungsten and xenon lamps | ||
monitor | 10.1 inch capacitive touch screen | ||
Operating system and language | Android/Chinese and English automatic switching | ||
Data output | USB1.1, Email, USB flash drive, WiFi | ||
Data output format | CSV,PDF | ||
RAM | 32GB(Maximum expandable to 256GB) | ||
Power range | AC90~250V/ 50~60Hz | ||
Size L × W × H) mm | 650×510×210 | ||
weight | 29kg | 30kg |
Q1: Why Choose NANBEI ?
(1).Professional manufacturer with more than 13 years experience
(2).Exported to more than 97% Countries
(3).Turnkey Solution is no problem
Q2:OEM,ODM acceptable or not?
Absolutely Yes
Q3:What's kind of Payment terms for customer choosing?
T/T ,Western Union, Money Gram , Credit Card, Paypal , L/C ...
Q4:Can we visit your factory online?
Absolutely no problem
Q5:Can online video inspection before shipment?
Absolutely no problem
Q6: what's the MOQ ? Sample order is OK?
MOQ:1 set, sample order is no problem
Q7:What's kind of shipment for customer choosing?
Usually ship by sea, by air, by international express.
We can also provide reasonable solutions according to your transportation requirements
Q8:How to ensure product quality and after-sales service?
We have CE, ISO quality certificate, and SGS authentication.
After-sale service:
1.Warranty:1 year
2.We supply free part for quality problem in warranty
3.Long life technical support and service